H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one mri ultra slimport implantable port was returned for evaluation.Gross visual, microscopic visual and functional evaluations were performed on the returned device.During functional evaluation, pressure was felt upon infusing the port and aspiration was unsuccessful.However, the investigation is inconclusive for the reported difficult to flush as only partial device was returned for evaluation and the exact circumstances at the time of the reported event cannot be verified.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 05/2025), g3.H11: g1, h6 (method, result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|