• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number SN6AT4
Medical Device Problem Code Application Program Problem: Power Calculation Error (1473)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported following an intraocular lens implant procedure, the calculation of the lens was found out to be calculated in an erroneous way.The surgeon performed a secondary surgery to rotate the lens, but the haptic was stuck very strong and the surgeon was not able to pull out the lens.It was discussed with the patient but the patient decided that she would preferred an external aid to correct the residual astigmatism (the one she had been using before surgery: toric contact lenses) and since it was resolved.The patient was satisfied with a toric contact lens after the surgery of the 1st eye.Additional information has been requested.
 
Additional Manufacturer Narrative
The product was not returned for analysis.The root cause for the reported complaint appears to be a coincidental event, that was not related to product, but to calculation issue.As it is stated in the file, that surgeon made the calculation of the lens in an erroneous way.A malfunction has not been indicated or information provided, that the lens was the cause of the event.Based on this information, the product did not cause/contribute to the event.Based on the results from the product history record, the products met release criteria.There have been no other complaints for this lot.Investigation has been completed, based on current information.No further action is warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14285430
Report Number9612169-2022-00202
Device Sequence Number5198864
Product Code MJP
UDI-Device Identifier00380652250467
UDI-Public00380652250467
Combination Product (Y/N)N
Initial Reporter CountryAR
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberSN6AT4
Device Catalogue NumberSN6AT4.220
Device Lot Number21259171
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/07/2022
Supplement Date Received by Manufacturer05/29/2022
Initial Report FDA Received Date05/05/2022
Supplement Report FDA Received Date06/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-