| Model Number |
C12787 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One quantien was received for analysis.Visual inspection found all labeling and input/output connectors to be free of physical damage.The unit was found to be coated in an unknown waxy, yellow material.The device was plugged in and smoke immediately began coming from the power supply brick.The unit was unplugged and the power supply brick was opened.Signs of a thermal event were identified at the input port on the power supply brick.The device was opened and the power supply was bypassed with a known good power supply.The unit powered on successfully when attempted.The reported complaint was confirmed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received and the investigation performed, the cause for the reported event was isolated to the power supply input port.
|
| |
|
Event or Problem Description
|
|
This report is to advise of a thermal event noted during analysis.
|
| |
|