| Model Number |
52-1403 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pulmonary Embolism (1498); Hypersensitivity/Allergic reaction (1907); Pain (1994); Osteolysis (2377)
|
| Date of Event |
01/24/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Asr litigation complaint received ad 31 jan 2022.Litigation records alleges pain, injury economic loss, punitive damages, excessive levels of chromium and cobalt.Doi: (b)(6) 2008.Dor: (b)(6) 2020, left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: a2, a3a, a4, b5, b6, b7, d3 (manufacturer email), h6 (health effect - clinical code), d4 (primary udi).Corrected: a1.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Medical records received.Revision left total hip arthroplasty was performed due to the severity of osteolysis and risk of periprosthetic fracture and protrusio.Intraoperatively, the hip was dislocated and the femoral head disimpacted.The acetabular component was malpositioned with excessive abduction and anteversion, contributing to an adverse reaction to metal debris.Significant eccentric wear of the metal liner and acetabular cup was identified.Litigation records ad 09 feb 2026.Additional diagnosis or condition experienced found, pseudotumor.Doi: (b)(6) 2008.Dor: (b)(6) 2020.Left hip.
|
| |
|
Search Alerts/Recalls
|