Additional Manufacturer Narrative
|
|
Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an unknown procedure.The liquid solution from the norian syringe would not inject into the rotary pouch injection port, instead spilling out all over the mixer and table, even though care was taken to ensure the syringe was securely attached to the port.The procedure was successfully completed with a 3 minutes of surgical delay noted.No fragments were generated.No patient outcome was reported.This report involves one (1) norian drillable inject 10cc-sterile.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H4 device history: part # 07.704.010s, synthes lot # ds7007689, release to warehouse date:12.01.2021, supplier: (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6 investigation summary: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the norian drillable inject 10cc-sterile had a cracked luer valve fitting.No other damage or defects were visible.A dimensional inspection was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the norian drillable inject 10cc-sterile would contribute to the complained device issue.During the investigation it was found that nr-0173166 was issued for this specific defect of luer lock cracking.A scar (supplier escalation) and nc has been assigned for the luer lock cracking.The pouch has been monitored by dsm biomedical per scar-000490 for luer lock cracking.Dsm¿s pouch supplier has identified an issue with their luer supplier.An investigation was conducted and documented device history: part # 07.704.010s.Synthes lot # ds7007689.Supplier lot # ds7007689.Release to warehouse date:12.01.2021.Supplier: dsm biomedical.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|