• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION COYOTE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION COYOTE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24705
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that balloon rupture occurred.Vascular access was obtained the via the right femoral artery.The 95% stenosed target lesion was located in the mildly tortuous and moderately calcified right anterior tibial artery.After inserting a non-boston scientific (bsc) introducer sheath, a non-bsc guidewire was used to cross the lesion.A 2.5mm x 150mm x 150cm, otw coyote balloon catheter was advanced for dilatation.However, during the first inflation at 10 atmospheres for 10 seconds, the balloon ruptured and a pinhole was noted in the middle part.The device was removed and the procedure was completed with another of same device.No patient complications were reported and the patient was in good condition post procedure.
 
Additional Manufacturer Narrative
E1- initial reporter city: (b)(6).Device evaluated by mfr.: returned product consisted of a coyote balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a pinhole 8mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that balloon rupture occurred.Vascular access was obtained the via the right femoral artery.The 95% stenosed target lesion was located in the mildly tortuous and moderately calcified right anterior tibial artery.After inserting a non-boston scientific (bsc) introducer sheath, a non-bsc guidewire was used to cross the lesion.A 2.5mm x 150mm x 150cm, otw coyote balloon catheter was advanced for dilatation.However, during the first inflation at 10 atmospheres for 10 seconds, the balloon ruptured and a pinhole was noted in the middle part.The device was removed and the procedure was completed with another of same device.No patient complications were reported and the patient was in good condition post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COYOTE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14287126
Report Number2134265-2022-05183
Device Sequence Number1559712
Product Code DQY
UDI-Device Identifier08714729796534
UDI-Public08714729796534
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111295
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/26/2025
Device Model Number24705
Device Catalogue Number24705
Device Lot Number0028756393
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/02/2022
Supplement Date Received by Manufacturer06/04/2022
Initial Report FDA Received Date05/05/2022
Supplement Report FDA Received Date06/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE: JUPITER NS15; GUIDEWIRE: JUPITER NS15; INFLATION DEVICE: EVEREST; INFLATION DEVICE: EVEREST; INTRODUCER SHEATH: HELLO ONE 5F; INTRODUCER SHEATH: HELLO ONE 5F; MICRO CATHETER: CORSIR PV; MICRO CATHETER: CORSIR PV
Patient SexUnknown
-
-