| Model Number |
1365-28-230 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient was revised due to the dislocation of head and liner and were replaced.New 28 head 958840 broke when removing.Unknown if any loosening.Unknown if any surgical delay.Doi: (b)(6) 2020, dor: (b)(6) 2022, affected side: left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: a1, a2 (age), d4 (expiration date) and h4.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d4 (lot).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation was performed for the finished device 136528230, lot 9588540, and no non-conformances / manufacturing irregularities were identified.The component properties and the microstructure as obtained from the quality documents fulfil the requirement as specified at the time of production.There is no indication of any pre-existing material defect.
|
| |
|