• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACCUMAX; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACCUMAX; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M0068404030
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that an accumax laser fiber was used in a percutaneous nephrolithotomy procedure for a lithiasis renal treatment in the kidney performed on (b)(6) 2022.The laser unit used was an accutech 60 watts laser console.During procedure, the laser fiber presented a lot of fragility and broke.No fiber fragments fell inside the patient.The procedure was completed with a different type of device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: medical device problem code a0401 captures the reportable event of fiber break.Block h10: investigation results the returned accumax laser fiber was analyzed, and a visual evaluation noted that the laser fiber was returned in a partial piece.The partial fiber piece measured 127 cm from the end of the connector towards the distal tip.There was no exposed glass tip present, this part of the fiber was not returned.When connected to the laser console, the aim beam was visible at the end of the fractured part.A functional evaluation revealed that the light from the sma connector was bright and round, indicating no fracture within the connector.No other problems with the device were noted.The reported event of the accumax laser fiber broke was confirmed.It is likely that procedural handling of the device during set-up or use contributed to the fiber body break.Review and analysis of all available information indicated that the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that an accumax laser fiber was used in a percutaneous nephrolithotomy procedure for a lithiasis renal treatment in the kidney performed on (b)(6) 2022.The laser unit used was an accutech 60 watts laser console.During procedure, the laser fiber presented a lot of fragility and broke.No fiber fragments fell inside the patient.The procedure was completed with a different type of device.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUMAX
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
COHERENT INCORPORATED
5100 patrick henry drive
santa clara CA 95054
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14287863
Report Number3005099803-2022-02427
Device Sequence Number5408194
Product Code GEX
UDI-Device Identifier08714729784609
UDI-Public08714729784609
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K082928
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/22/2024
Device Model NumberM0068404030
Device Catalogue Number840-403
Device Lot Number0000010975
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/11/2022
Supplement Date Received by Manufacturer06/08/2022
Initial Report FDA Received Date05/05/2022
Supplement Report FDA Received Date06/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-