MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2021
Event Type  malfunction  
Event Description
After a completed case, the sheath subsequently tested out of specification per the manufacturers investigation.
 
Manufacturer Narrative
Product event summary: the 4fc12 sheath with lot number 0010150268 was returned and analyzed.Visual inspection of the shaft area was performed.The inspection identified shaft kink/twists at approximately five inches, seven inches, and nine inches from the tip.In conclusion, the sheath failed the returned product inspection due to shaft kinks.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14288050
MDR Text Key290792313
Report Number3002648230-2022-00190
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/02/2022
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0010150268
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2022
Date Manufacturer Received04/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-