| Model Number |
N/A |
| Medical Device Problem Codes |
Failure to Advance (2524); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported a nester platinum embolization coil failed to deploy in an (b)(6) male patient during a false lumen embolization procedure.The physician attempted to push the coil through the catheter, but it became stuck in the tip of the catheter.The catheter was removed and another was used to complete the procedure successfully.Photos provided by the customer showed the coil lodged in the distal tip of the catheter.No adverse effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Pma/510(k) #: preamendment.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
| |
|
Event or Problem Description
|
|
On 11may2022, the device was returned with the coil lodged in the distal tip of the catheter.A second coil was also returned.It's currently unknown why the second coil was returned.
|
| |
|
Additional Manufacturer Narrative
|
|
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Additional information provided 05may2022 confirmed the device was flushed prior to introduction and an 0.035 competitor wire guide was used.
|
| |
|
Additional Manufacturer Narrative
|
|
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Two coils were returned to cook for investigation.It was confirmed on 08jun2022 that only one coil rpn: mwce-35-14-12-nester was stuck at the tip of the catheter.The user is unable to recall why two coils were returned to the manufacturer and confirmed the complaint was only for one coil.
|
| |
|
Event or Problem Description
|
|
No additional information regarding patient and/or event details has been received since the previous medwatch report was sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation ¿ evaluation (b)(6) hospital (taiwan) informed cook on 21apr2022 of an incident that occurred on the same date involving a nester platinum embolization coil (rpn: mwce-35-14-12-nester, lot 14199510).It was reported that when the user attempted to advance the coil through the catheter, it was difficult to advance and became stuck at the tip of the catheter.The procedure was then finished successfully using a new device, and there were no adverse effects reported to the patient due to this occurrence.Reviews of the documentation, including the complaint history, device history record (dhr), instructions for use (ifu) and quality control procedures, as well as a visual inspection of the returned device, were conducted during the investigation.One catheter was returned with the coil stuck at the tip of the catheter.After removal from the catheter, the coil was found to be separated and elongated; however, this is believed to have occurred following removal from the catheter during the investigation, due to the coil being lodged at the tip.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are currently in place to identify this failure mode prior to distribution.A review of the device history record (dhr) for lot 14199510 found one possibly related nonconformance for an incorrect curl diameter, in which all nonconforming devices were scrapped following 100% inspection.It should be noted that there were no other complaints associated with this lot number.Cook also reviewed product labeling.The current instructions for use [t_ce_nec_rev4] state the following: "instructions for use 2.Firmly grasp the loading cartridge between thumb and forefinger.Introduce the metal end of the loading cartridge into the base of the catheter hub.Lock loading cartridge onto catheter hub by turning luer lock adapter lock adapter clockwise.3.Maintaining position of the cartridge, advance the stiff portion of the wire guide into the loading cannula.Push the coil into the first 20 to 30 cm of the angiographic catheter.Remove the wire guide and loading cartridge.4.With the flexible tip of the wire guide, advance the embolization coil to the tip of the catheter.Verify position of the angiographic catheter prior to deployment.5.Deploy the coil by advancing the wire guide past the tip of the catheter.How supplied upon removal from package, inspect the product to ensure no damage has occurred." evidence gathered upon review of the dmr, product labeling, and dhr, does not indicate the device was manufactured out of specification.There is no evidence of nonconforming material in house or in the field.Based on the information provided, examination of the returned product and the results of our investigation, it was concluded that a component failure unrelated to manufacturing or design deficiencies contributed to the event.The appropriate personnel have been notified.Per the risk assessment no further action is required.We will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|