• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA¿ XT CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA¿ XT CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTBA1D4
Medical Device Problem Codes Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 04/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient received a potentially inappropriate shock for an atrial arrhythmia with rapid ventricular response which was detected by the cardiac resynchronization therapy defibrillator (crt-d) as ventricular fibrillation (vf).It was also reported that there was occasional undersensed atrial beats noted during episodes.The crt-d and right atrial (ra) lead remain in use. no further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA¿ XT CRT-D
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key14288403
Report Number3004209178-2022-05688
Device Sequence Number11152768
Product Code NIK
UDI-Device Identifier00643169929302
UDI-Public00643169929302
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/28/2018
Device Model NumberDTBA1D4
Device Catalogue NumberDTBA1D4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2022
Initial Report FDA Received Date05/05/2022
Date Device Manufactured05/09/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6947M55 LEAD, 419688 LEAD, 5568-45 LEAD.
Outcome Attributed to Adverse Event Life Threatening;
Patient SexFemale
Patient Weight64 KG
-
-