• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES ARCHITECT TACROLIMUS REAGENT KIT; ENZYME IMMUNOASSAY, TRACROLIMUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES ARCHITECT TACROLIMUS REAGENT KIT; ENZYME IMMUNOASSAY, TRACROLIMUS Back to Search Results
Model Number 1L77-25
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer questioned architect tacrolimus results for one patient.The patient is a 15 year-old female, sid (b)(6) result of 2.98 pg/ml (reference range: 5-20) was questioned by the physician as skewed low.The patient was originally tested at another laboratory using a different platform, with results all within the reference range.No impact to patient management was reported.
 
Additional Manufacturer Narrative
The evaluation of complaint data for the product and likely cause architect tacrolimus reagent lot 33444fp00 identified normal complaint activity and there are no trends for the product related to patient results.No customer returns were available for evaluation.Inhouse testing determined that the accuracy performance is not negatively impacted.Manufacturing documentation for the likely cause lot did not identify any issues associated with the complaint issue.The performance of the likely cause lot was investigated by completing a review for non-conformances, potential non-conformances and deviations related to the likely cause lot.This review did not identify any non-conformances, potential non-conformances, or deviations.A review of labeling concluded that the issue is sufficiently addressed.Based on our investigation no systemic issue or product deficiency of the architect tacrolimus reagent lot 33444fp00 was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT TACROLIMUS REAGENT KIT
Common Device Name
ENZYME IMMUNOASSAY, TRACROLIMUS
Manufacturer (Section D)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064
Manufacturer (Section G)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key14288706
Report Number1415939-2022-00034
Device Sequence Number20185348
Product Code MLM
UDI-Device Identifier00380740001377
UDI-Public00380740001377
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K070820
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/07/2023
Device Model Number1L77-25
Device Catalogue Number01L77-25
Device Lot Number33444FP00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/21/2022
Supplement Date Received by Manufacturer06/07/2022
Initial Report FDA Received Date05/05/2022
Supplement Report FDA Received Date06/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/02/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC I2000SR INST, 03M74-02, ISR64078; ARC I2000SR INST, 03M74-02, ISR64078
Patient Age15 YR
Patient SexFemale
-
-