Catalog Number 03.111.022
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during inspection of back up instruments on an unknown date, the depth gauge for 2.0mm and 2.4mm screws had a loose needle, two handle with mini quick coupling had issues locking driver shaft, two hold-sl ø2 f/compression/distraction dev did not lock wires, compression/distraction instrument turn knob kept popping off, comp/distraction instr for ulna osteo had a wire stuck inside the device, and a t-handle with quick coupling had a conical extraction screw stuck inside.There was no hospital involvement.This report is for a compression/distraction holding sleeve/2.0mm k-wire.This is report 3 of 6 for (b)(4).
 
Additional Manufacturer Narrative
Date of event is an unknown date in 2022.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPRESSION/DISTRACTION HOLDING SLEEVE/2.0MM K-WIRE
Common Device Name
INSTRUMENT, COMPRESSION
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14288834
Report Number2939274-2022-01616
Device Sequence Number1519171
Product Code HWN
UDI-Device Identifier07611819375826
UDI-Public(01)07611819375826
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.111.022
Device Lot Number2550553
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/04/2022
Initial Report FDA Received Date05/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
COMP/DISTRACTION INSTR FOR ULNA OSTEO; COMPRESSION/DISTRACTION INSTRUMENT; DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; HANDLE WITH MINI QUICK COUPLING; HANDLE WITH MINI QUICK COUPLING; HOLD-SL Ø2 F/COMPRESSION/DISTRACTION DEV; T-HANDLE WITH QUICK COUPLING
Patient SexUnknown
-
-