Model Number VCP305H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Did two sutures break during the procedure? yes/no.There was another suture broken or did the needle separated from the suture during the procedure? what is the lot number? was the needle removed from the patient? any additional tissue dissection damage due to searching for needle? any other patient consequences? if yes, what medical/surgical intervention was provided.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Trade name - irgacare®, active ingredient(s) ¿ triclosan, dosage form ¿ suture/solid/parenteral, strength ¿ = 275 ug/m.Event related to mw # 2210968-2022-03337, mw # 2210968-2022-03338.This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that a patient underwent a kidney surgery procedure on (b)(6) 2022 and suture was used.The suture broke.A like device was used to complete the procedure.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 6/14/2022.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: twenty packets of product code vcp305h were returned to ethicon for analysis with the packaging closed.Upon initial inspection to the samples no external damages were observed on the packets.In order to evaluate the conditions of the returned samples, thirteen packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand to detect any issue related to damaged or breakage suture and no defects were observed during evaluation.A functional test was performed using an instron and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: thirty-five packets of product code vcp305h were returned to ethicon for analysis with the packaging closed.Upon initial inspection to the samples no external damages were observed on the packets.In order to evaluate the conditions of the returned samples, thirteen packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand to detect any issue related to damaged or breakage suture and no defects were observed during evaluation.A functional test was performed using an instron and the tensile strength force was above the minimum requirement.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Trade name - irgacare®; active ingredient(s) ¿ triclosan; dosage form ¿ suture/solid/parenteral; strength ¿ = 275 g/m.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) date sent to the fda: 5/24/2022 this is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, the following was obtained: there were two suture broken during the procedure? yes/no - yes there was another suture broken or did the needle separated from the suture during the procedure? -the suture broke and the needle detached from the suture where the suture connected to the needle.What is the lot number? i provided the customer service representative with the lot number.Was the needle removed from the patient? yes any additional tissue dissection damage due to searching for needle? no any other patient consequences? if yes, what medical/surgical intervention was provided - no are any samples available of evaluation? please provide facility contact person¿s name, mailing address and telephone.A shipper kit will be sent to help return the product to us.- the surgeon did not keep the defective product.I recently sent back all of the additional suture that was left in the 2 boxes.Trade name - irgacare® active ingredient(s) ¿ triclosan dosage form ¿ suture/solid/parenteral strength ¿ = 275 ¿g/m.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VICRYL PLUS SUTURE 27"(70CM) 3-0 VIO
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key14288895
Report Number2210968-2022-03335
Device Sequence Number5408804
Product Code GAM
UDI-Device Identifier10705031052130
UDI-Public10705031052130
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
K132580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model NumberVCP305H
Device Catalogue NumberVCP305H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2022
Supplement Date Received by Manufacturer05/11/2022
05/23/2022
Initial Report FDA Received Date05/05/2022
Supplement Report FDA Received Date05/24/2022
06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-