• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US TRACKER SLOPESELECT ARM *EA; BONENAIL GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US TRACKER SLOPESELECT ARM *EA; BONENAIL GUIDEWIRE Back to Search Results
Model Number 219455
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
This is report 2 of 2 for (b)(4).It was reported by the sales rep that during an unknown procedure (b)(6) 2022, it was observed that the tracker bone clamp device was not working properly.According to the report, the clamp device would not stay clamped by itself.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.This complaint involves two (2) devices.
 
Additional Manufacturer Narrative
Additional narrative: udi: (b)(4).The lot number was unknown.The lot number was unknown; therefore, the expiration date, manufacturing site name and device manufacture date were unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary : the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.Given that no lot number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
BONENAIL GUIDEWIRE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14290793
Report Number1221934-2022-01310
Device Sequence Number13835958
Product Code NBH
UDI-Device Identifier10886705006571
UDI-Public10886705006571
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number219455
Device Catalogue Number219455
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2022
Supplement Date Received by Manufacturer05/26/2022
Initial Report FDA Received Date05/05/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-