• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTUA; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTUA; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number V272
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product has been received for analysis.Upon completion of the failure analysis of the complaint product, if there is any further relevant information from that review, a supplemental report will be filed.
 
Event or Problem Description
It was reported that there was a problem with the batteries of this cardiac resynchronization therapy pacemaker (crt-p) device.Subsequently, the patient underwent a replacement procedure, and this device was explanted and replaced.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
The product has been received for analysis.Upon completion of the failure analysis of the complaint product, if there is any further relevant information from that review, a supplemental report will be filed.The returned cardiac resynchronization therapy pacemaker (crt-p) was analyzed and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, defibrillation, and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
 
Event or Problem Description
It was reported that there was a problem with the batteries of this cardiac resynchronization therapy pacemaker (crt-p) device.Subsequently, the patient underwent a replacement procedure, and this device was explanted and replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTUA
Common Device Name
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14291083
Report Number2124215-2022-14367
Device Sequence Number16527153
Product Code NKE
UDI-Device Identifier00802526536663
UDI-Public00802526536663
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P030005/S092
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/12/2017
Device Model NumberV272
Device Catalogue NumberV272
Device Lot Number605049
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/09/2022
Supplement Date Received by Manufacturer10/07/2022
Initial Report FDA Received Date05/05/2022
Supplement Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/12/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age77 YR
Patient SexFemale
-
-