Model Number M00583210
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a truetome dreamwire 44 was used in the papilla of vater during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the cutting wire of the truetome dreamwire 44 broke while cutting.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Additional Manufacturer Narrative
Block d4, h4: the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Block h6 (device codes): medical device problem code a0401 captures the reportable event of cutting wire broken.Block h10: the returned truetome dreamwire 44 was analyzed, and a visual evaluation noted that the cutting wire was broken from the proximal pierced hole.The device was observed under magnification and the cutting wire was broken, bent, and blackened.It was also found that the distal pierced hole was torn.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken and blackened.The cutting wire being blackened indicates that the device was energized.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can kink the cutting wire and tear the distal pierce hole.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device was unable to be performed as the lot number is unknown.However, a ship history review was performed to identify the most probable lots, and a device history review (dhr) on the most probable lots did not identify any anomalies or deviations within manufacturing/service processes that could have contributed to the event.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a truetome dreamwire 44 was used in the papilla of vater during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the cutting wire of the truetome dreamwire 44 broke while cutting.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETOME DREAMWIRE 44
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14291342
Report Number3005099803-2022-02471
Device Sequence Number1532401
Product Code KNS
UDI-Device Identifier08714729858751
UDI-Public08714729858751
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K122203
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00583210
Device Catalogue Number8321
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer05/17/2022
Initial Report FDA Received Date05/05/2022
Supplement Report FDA Received Date06/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-