| Model Number |
N/A |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
04/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Report source: (b)(6).Concomitant medical products: catalog number: 110010241, lot number: 6297939, brand name: g7 osseoti46mm ltd b.Catalog number:110024460, lot number: 707400, brand name: g7 dm metal liner 36mm b.Catalog number: 00771301000, lot number: 63958381, brand name: ml taper kinectiv size10.Catalog number: 00784801200, lot number: 63958196, brand name: modular neck size e.Catalog number: 00801802202, lot number: 63904959, brand name: cocr head +0x22mm.Catalog number: 010000991, lot number: 602010, brand name: e1aa brg 36mm id22mm.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was revised due to hip dislocation and disassociation of the head and bearing approximately 4 years post implantation.Attempts have been made and additional information on the reported event is unavailable.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, h1, h2, h3, h6, h10 h6: component code: mechanical - g04 head visual examination of the provided pictures identified the bearing/head and stem with the bearing showing damage to the outside radius, it is unknown if that damage was caused invivo or during extraction of the device medical records/radiographs were provided and reviewed by a health care professional.A review of the available records identified findings of the reported issues: interval dislocation of the right hip arthroplasty reported event was confirmed by review of x-rays provided.Review of the device history records identified no deviations or anomalies during manufacturing.Root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|