• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSION ORTHOPEDICS, INC. RSS GLENOID BASEPLATE IMPACTOR-S; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSION ORTHOPEDICS, INC. RSS GLENOID BASEPLATE IMPACTOR-S; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number IMP-0961-064-S
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that the plastic tips of an rss glenoid baseplate impactor-s have worn down and broken off after extensive use.No case was reported; therefore, there was no patient involvement.
 
Additional Manufacturer Narrative
The product was received at the aus site for evaluation.The complaint was confirmed.The following investigative actions were performed.Refer to the respective tasks for further detail.Complaint history review: the complaint history review identified similar reported events for this device.No previous events or issues were identified for the same product lot.No adverse trend was identified.Future complaints for this failure mode will continue to be monitored and investigated as required.Risk management review (device): identified no issues which could have caused or contributed to the reported event.The failure mode was previously identified by the risk management file and the anticipated risk level is acceptable.Dhr/batch record review: identified no issues or manufacturing abnormalities which could have caused or contributed to the reported event.This review determined that the device met manufacturing specifications upon release for distribution.Capa/nc/pra/hhe/field action review: identified no previous events or issues which could have caused or contributed to the reported event.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.Device labeling/ifu review: identified no issues which could have caused or contributed to the reported event.This review determined that the device met labeling requirements upon release for distribution.Product prints/specifications/procedure review: identified no issues which could have caused or contributed to the reported event.This review determined that the device met product specifications upon release for distribution.Visual inspection: verified that the most probable cause for the failure was wear and damage due to normal use.Evaluation of the returned device found that the black plastic (radel) tip of the instrument was broken off and missing.The complaint alleges no injury to the patient nor a delay during surgery.It was determined that the device likely contributed to the reported event as the most probable cause of the failure was wear and damage due to normal use.The rss glenoid baseplate impactor is a reusable instrument expected to be used across multiple surgeries.Further, its intended use is to absorb forces exerted on the glenoid baseplate, an implantable component of the reverse shoulder system, during impaction in order to prevent damage to the implant.Therefore, the plastic tip of the device may fail after prolonged use or if subjected to excessive force, such as during impaction.This probable cause is supported by similar previous complaints.Based on this investigation, the need for corrective action is not indicated as no non-conformances or product deficiencies were identified and the risk level is acceptable.If additional information is later received, the complaint may be reopened.No further investigation is warranted for this complaint, though future complaints will be monitored and investigated as required.This investigation can be closed.H8: usage of device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSS GLENOID BASEPLATE IMPACTOR-S
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer (Section G)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14291988
Report Number3002788818-2022-00005
Device Sequence Number1595998
Product Code LXH
UDI-Device Identifier10381780467786
UDI-Public10381780467786
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIMP-0961-064-S
Device Catalogue NumberIMP-0961-064-S
Device Lot Number245623-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2022
Supplement Date Received by Manufacturer05/27/2022
Initial Report FDA Received Date05/05/2022
Supplement Report FDA Received Date05/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-