It was reported that, after a tka surgery had been performed on (b)(6) 2010, the journ art ins bcs std 3-4 rt 9 broke.The patient is in pain and impacted by the complication.A partial removal of the broken part was performed under arthroscopy in early (b)(6) 2022.A revision surgery date is to be set to replace the insert.The current health status of the patient is unknown.
|
|
H10: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that without the requested clinical documentation, the root cause of the reported event could not be fully assessed.The patient impact beyond the reported ¿patient very painful and impacted by the complication¿, post breakage, arthroscopy removal of the post-fragment, and planned insert revision could not be determined.No further medical assessment could be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that the implant can break or become damaged as a result of strenuous activity or trauma, and has a finite expected service life and may need to be replaced in the future.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to traumatic injury or surgical technique.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: (b)(4).
|