Model Number 66024008
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
During set-up/inspection of treatment, when a iv3000 1 hand 10x12cm pouch was opened, it was confirmed there was a dust-like foreign substance in it.Treatment was completed, without any delay, with a backup.No injury or other complications were reported.
 
Additional Manufacturer Narrative
The device was not returned for analysis.No photographs were provided in support of this complaint.We have not been able to confirm a relationship between the event and the device or identify a definitive root cause.It was reported during set-up/inspection of treatment, when a iv3000 1 hand 10x12cm pouch was opened, it was confirmed there was a dust-like foreign substance in it.Probable cause is that the dust is too small and too light in color, the equipment cannot be detected and rejected.The batch records were reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of the finished product specification.There was also no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.A complaint history review revealed a similar instances in the last three years.A review of prior escalation actions found no actions applicable to the scope of this case.A risk management review concluded that without further information the alleged failure mode can not be directly linked to a specific failure mode within the risk file although it does contain several failure modes that can result in contamination, of which have a worst case risk rating of medium risk level and is still adequate.This investigation is now complete with no further action deemed necessary.Smith and nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 1 HAND 10X12CM CTN 50
Common Device Name
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
101 hessle road
hull east riding of yorkshire HU3 2-BN
UK   HU3 2BN
5123913905
MDR Report Key14292285
Report Number8043484-2022-00152
Device Sequence Number1596004
Product Code KGX
UDI-Device Identifier05000223417406
UDI-Public05000223417406
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2024
Device Model Number66024008
Device Catalogue Number4008
Device Lot Number202105
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2022
Supplement Date Received by Manufacturer08/19/2022
Initial Report FDA Received Date05/05/2022
Supplement Report FDA Received Date08/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-