Model Number 801-36545M
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device remains implanted.
 
Event or Problem Description
A company representative reported an axial slip of a mesa deformity uniplanar screw.Revision surgery has not been scheduled.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.Device history records could not be reviewed without either the device or a valid lot number.Complaint history was reviewed for the reported catalog and similar complaints for were identified, originating from the same surgeon.Although post-operative x-rays were provided, the images are taken from different angles and there appears to be no discernable slip.The patient is fixated on alternating levels.Similar construct design x-rays from the same surgeon for the same failure mode were reviewed by a consultant spine health care provider on (b)(6) 2021.The physician believes that the construct design with difficult deformities was the most likely contributing factor to the event.The operating surgeon used a low density screw construct which ended on the convex curve of the spinal deformity.The low screw density and short length of construct, along with possible loss of correction, may not have been enough to counter the force of the spine trying to revert to its original position.This may have resulted in the eventual failure at the distal most screw (where biomechanical loading is most likely greater) leading to the rod slipping from the screw.It cannot be confirmed if an axial slip occurred from the images provided and a root cause cannot be established.If additional information is received or the device is returned for evaluation, the investigation will be reopened and updated.
 
Event or Problem Description
A company representative reported an axial slip of a mesa deformity uniplanar screw.Revision surgery has not been scheduled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEFORMITY, UNIPLANAR SCREW; SIZE Ø6.5X45 MM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key14292337
Report Number3004774118-2022-00177
Device Sequence Number19251374
Product Code NKB
UDI-Device Identifier10888857140301
UDI-Public10888857140301
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K141873
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number801-36545M
Device Catalogue Number801-36545M
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/20/2022
Supplement Date Received by Manufacturer07/16/2022
Initial Report FDA Received Date05/05/2022
Supplement Report FDA Received Date08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-