| Model Number |
APB-3-8-3D-ES |
| Medical Device Problem Codes |
Break (1069); Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that there was a breakage observed to the coil when the bag was opened.The coil was not used in the patient's body.There was no surgical or medicinal intervention required.It was unknown if there was deformation or damage to the delivery pusher.A continuous flush was administered during the procedure.The devices were prepared as indicated in the ifu.The patient was undergoing surgery for treatment of a saccular, ruptured aneurysm in the left ic-pc with a max diameter of 5mm and a 3mm neck diameter.It was noted the patient's blood flow was normal and vessel tortuosity was normal.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: as found condition: the axium prime coil was returned for analysis within a shipping box; within a plastic bag; within an opened axium prime outer carton; and within a dispenser coil.Visual inspection/damage location details: the axium prime implant coil was returned within the introducer sheath.No bent or kink were found with axium prime pushwire.The axium prime implant coil was found to be damaged and stretched within introducer sheath.No damages or irregularities were found with the introducer sheath.No other anomalies were observed.Testing/analysis: the axium prime implant coil was able to be pushed out from introducer sheath with resistance.Conclusion: based on the device analysis and reported information, the customer¿s report of ¿coil kink/damage ¿was confirmed.However, the customer¿s report of ¿coil separation/break¿ could not be confirmed as no separation was found with the returned axium prime coil.The axium prime implant coil was found to be damaged and stretched within introducer sheath.The investigation determined that this event was similar to an event that had already been investigated, and another investigation is not necessary.Based on the formal investigation, it is possible the damage occurred during production or upon opening the package and attempting to remove the product or after removing it from the package.No corrective action was required; however, as a conservative approach a quality alert was then issued to improve operator awareness for all products that are currently being manufactured internally at all medtronic sites that manufacture neurovascular product.The review of lot history records shows that the finished device has met all manufacturing requirements and specifications during final assembly and quality inspection.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported there was no damage or evidence of tampering to device packaging.It was unclear if the proximal end of the pushwire was secured in the guidewire clip (black rubber stopper) when the package was opened.According to the rep, the doctor said he thought that it was.The damage was noticed upon opening the package.Preparation was performed prior to the damage being seen.The first coil was already used.The issue was found when the package of the second coil was opened.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|