| Model Number |
PXB35-07-17-135 |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician was attempting to use a visi pro balloon expandable stent along with non-medtronic 6fr sheath and 0.014" guidewire during procedure to treat a moderately calcified lesion in the left common iliac artery (cia) and tibial/popliteal trunk (tpt).The vesseldiameter and lesion length are 7mm and approximately 70mm respectively.The vessel was moderately tortuous.No embolic protection was used.There was no damage noted to packaging.There was no issue noted when removing device from hoop/tray.The device was prepped per ifu with no issues identified.Stent dislodgement occurred during delivery to/at lesion.The vessel was pre dilated with a 6mm pre dilation device.The device passed through a previously deployed stent.There was resistance encountered when advancing the device with excessive force used during delivery.It was reported that physician place a 7 x 57 visi pro in iliac artery via left radial approach and needed to extend the stent slightly.When physician was advancing the 7 x 17 stent it was getting stuck on the shelf of the 7 x 57 stent that was previously placed.The stent migrated distally into sfa and eventually to popliteal/tp trunk and had no catheters that would reach from radial and needed to call surgeon for removal of stent.Patient was taken to surgery to remove dis lodged stent.Stent migration was complication associated with this event.No further patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|