Model Number ETUF3214C102E
Medical Device Problem Codes Leak/Splash (1354); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Shock (2072); Rupture (2208); Insufficient Information (4580)
Date of Event 05/03/2022
Type of Reportable Event Death
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant ii aui was implanted during an endovascular procedure for the treatment of a ruptured 60mm abdominal aortic aneurysm.It was reported during the index procedure the aui was successfully implanted along with 2 endurant limbs as iliac extension to seal the distal end of the aui graft.A possible type ia endoleak was observed.Due to lack of inventory no further devices were available for implant.It was reported the patient did not recover and expired.Due to shock the ruptured abdominal aortic aneurysm.  as per the physician the cause of the type ia endoleak is undetermined.The cause of death was due to shock from the ruptured aaa.No additional clinical sequalae was provided and the patient has expired.
 
Additional Manufacturer Narrative
B5; additional information was received: it was reported the patient was initially stable, but became quickly unstable due to rupture requiring emergent intervention with initial balloon occlusion control.The mdt sales rep attempted to obtain more product but only had access to a single aui device and no aortic cuffs.The option of acquiring alternative devices from a facility several hours away was considered but the patient's condition was deemed too unstable.The patient had extremely tortuous anatomy, requiring 2 iliac extensions and partially covering the lt ra (neck was noted to be hostile, maybe 22mm diameter, short).The lt ci was embolized with plans for a fem-fem bypass.A type ia endoleak was noted on the intraop angio and patient remained unstable during the case and eventually expired.The physician felt all was done that could have been in this patient with a rupture and was not dissatisfied with medtronic.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II AUI STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14302874
Report Number9612164-2022-01682
Device Sequence Number5000819
Product Code MIH
UDI-Device Identifier00643169232617
UDI-Public00643169232617
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/18/2022
Device Model NumberETUF3214C102E
Device Catalogue NumberETUF3214C102E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2022
Supplement Date Received by Manufacturer05/13/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date06/09/2022
Date Device Manufactured05/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age83 YR
Patient SexFemale
-
-