• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 1; HIP CEMENTLESS STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 1; HIP CEMENTLESS STEM Back to Search Results
Model Number 01.12.021
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 04/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 06 april 2022: lot 152958: (b)(4) items manufactured and released on 07-oct-2015.Expiration date: 2020-09-12.No anomalies found related to the problem.To date, all items of the same lot have been already sold without any similar reported event since 2018.Clinical evaluation performed by medacta medical director: about 7 years after primary cementless tha in an elderly lady, extensive radiolucent lines are visible on the radiogrphic control around the stem.In absence of a silent infection, this is probably a case of aseptic loosening, a possible adverse event following tha, described in literature.Causes remain often undetermined.We see no discernible clues that could point to a specific cause for this adverse event.
 
Event or Problem Description
Revision surgery following stem loosening 6 years and 5 months after primary.The stem, liner, and head were revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 1
Common Device Name
HIP CEMENTLESS STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key14303478
Report Number3005180920-2022-00328
Device Sequence Number14664368
Product Code JDI
UDI-Device Identifier07630030802027
UDI-Public07630030802027
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K082792
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/12/2020
Device Model Number01.12.021
Device Catalogue Number01.12.021
Device Lot Number152958
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/06/2022
Initial Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/07/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age87 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-