Avanos medical inc.Received a single report that referenced two different incidences, which were associated with separate units, involving two different patients.This is the first of two reports.Refer to 2026095-2022-00046 for the second report.The physician reported, the external bolster on the percutaneous endoscopic gastrostomy (peg) was loose resulting in tube displacement.The elderly patient was transferred to long term care (ltc) following the procedure, the nursing staff reported leaking at the site.The patient returned to endoscopy (ed) and tube misplacement was noted; the patient passed a few days later of sepsis related complications.Additional information received 08apr2022 reported, after the tube placement, the bolster was noted to be looser than normal and the bolster was strengthened with a silk suture, the patient was discharged the following day.The patient was readmitted, through the emergency room (er), with a diagnosis of peritonitis.Upon examination in the er the bolster was noted to be far from the abdominal wall and there was leakage around the tube; it was additionally noted that the patient had a firm and tender abdomen.The nursing home staff reported that they did not use the tube, no reason was provided.The bolster was maintained tight to the skin however, the tube was not used during admission.The patient was treated with antibiotics and intensive care; the patient succumbed to secondary peritonitis, renal failure and sepsis.
|
|
The actual sample from the reported event was not returned for evaluation; however, one photo was provided where it was observed that the secure lock had a 24 engraved instead of 20.An investigation was performed for this incident and a manufacturing process was verified based on the photo provided by the customer.The reported event could be confirmed as reported: mixed product; the root cause is manufacturing related.All information reasonably known as of 08 nov 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint: (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|