Internal investigation into strattice lot sp100288 included a review of the reported information, review of the device history records, and a review of the complaint history records.The investigation resulted in no remarkable findings, including no other confirmed related complaints reported against the lot and no deviations or related non-conformances revealed during processing.The lot was terminally sterilized within the process parameters and met all qc release criteria.As of 4/27/2022, of the 221 devices released to finished goods for lot sp100288, 188 have been distributed with 141 reported as implanted.Based on our internal review with no remarkable findings, a relationship between the strattice and this event could not be determined.No further actions are required as a nonconformance could not be confirmed.
|
|
It was reported through a legal event that a (b)(6) male patient underwent hernia repair surgery on or about (b)(6) 2015.During the hernia repair surgery, the surgeon implanted a strattice mesh lot sp100151-096.After surgery, the patient returned to the hospital on or about (b)(6) 2016 and was diagnosed with a recurrent hernia.On or about (b)(6) 2016 the patient underwent a mesh revision operation.As part of the revision surgery, the surgeon implanted another strattice mesh, lot no.Sp100288-057.Following the revision surgery, the patient again returned to the hospital on (b)(6), 2016, where the patient was diagnoses with a recurrent incarcerated ventral hernia.On (b)(6) 2016, the surgeon implanted a third strattice mesh, lot no.Sp100503-132.This record is associated with the second device implanted on (b)(6), 2016, lot sp100288-057.As per the reported information, there is no complaint against the third device lot sp100503-132.
|