• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRATTICE LAPAROSCOPIC 16CM X 20CM; MESH, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRATTICE LAPAROSCOPIC 16CM X 20CM; MESH, SURGICAL Back to Search Results
Catalog Number 1620005
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hernia (2240)
Date of Event 06/09/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal investigation into strattice lot sp100288 included a review of the reported information, review of the device history records, and a review of the complaint history records.The investigation resulted in no remarkable findings, including no other confirmed related complaints reported against the lot and no deviations or related non-conformances revealed during processing.The lot was terminally sterilized within the process parameters and met all qc release criteria.As of 4/27/2022, of the 221 devices released to finished goods for lot sp100288, 188 have been distributed with 141 reported as implanted.Based on our internal review with no remarkable findings, a relationship between the strattice and this event could not be determined.No further actions are required as a nonconformance could not be confirmed.
 
Event or Problem Description
It was reported through a legal event that a (b)(6) male patient underwent hernia repair surgery on or about (b)(6) 2015.During the hernia repair surgery, the surgeon implanted a strattice mesh lot sp100151-096.After surgery, the patient returned to the hospital on or about (b)(6) 2016 and was diagnosed with a recurrent hernia.On or about (b)(6) 2016 the patient underwent a mesh revision operation.As part of the revision surgery, the surgeon implanted another strattice mesh, lot no.Sp100288-057.Following the revision surgery, the patient again returned to the hospital on (b)(6), 2016, where the patient was diagnoses with a recurrent incarcerated ventral hernia.On (b)(6) 2016, the surgeon implanted a third strattice mesh, lot no.Sp100503-132.This record is associated with the second device implanted on (b)(6), 2016, lot sp100288-057.As per the reported information, there is no complaint against the third device lot sp100503-132.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATTICE LAPAROSCOPIC 16CM X 20CM
Common Device Name
MESH, SURGICAL
Manufacturer Contact
christopher belle
1 millennium way
branchburg, NJ 08876
9089471470
MDR Report Key14305436
Report Number1000306051-2022-00088
Device Sequence Number4795117
Product Code FTM
UDI-Device Identifier00818410012885
UDI-Public00818410012885
Combination Product (Y/N)N
Initial Reporter StatePA
PMA/510(K) Number
K130817
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number1620005
Device Lot NumberSP100288
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2022
Initial Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NO INFORMATION REPORTED
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-