• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. BALLARD CLOSED SUCTION SYSTEMS FOR NEONATES/PEDIATRICS, ELBOW; VAP CLOSED SUCTION CATHETERS & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. BALLARD CLOSED SUCTION SYSTEMS FOR NEONATES/PEDIATRICS, ELBOW; VAP CLOSED SUCTION CATHETERS & ACCESSORIES Back to Search Results
Model Number 210
Medical Device Problem Code Suction Problem (2170)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the closed suction catheter (csc) did not suction during use.There was no reported injury.
 
Additional Manufacturer Narrative
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 04 may 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The device history record for lot 30154260 was reviewed and the product was produced according to product specifications.All information reasonably known as of 01 jun 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
Health effect - clinical code.Health effect -impact codes.The actual sample from the reported event was returned for evaluation.Visual examination of the device revealed the thumb valve was broken from the control body and could not be rotated into the unlocked position; the position of the thumb valve was in an opened position.The skives were visible on the distal end of the catheter and there was no blockage noted at the tip of the catheter.Additionally, there were no signs of damaged/ pinched/ curved tubing on the catheter.Testing of the sample revealed: the stem portion of the valve bounced back up when it was pressed and released multiple times; and suctioning occurred when the stem of the valve was attached to suctioning equipment (set at 120mmhg) and the distal end of the catheter was inserted in a beaker of water with blue dye.The reported event could be confirmed; however, no root cause was identified for the reported issue since no failure was detected during the evaluation of the returned device.The device history record for lot 30154260 was reviewed and the product was produced according to product specifications.All information reasonably known as of 29 sep 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLARD CLOSED SUCTION SYSTEMS FOR NEONATES/PEDIATRICS, ELBOW
Common Device Name
VAP CLOSED SUCTION CATHETERS & ACCESSORIES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S DE RL DE CV
carretera internacional
salida norte no. 1053
magdalena, cp
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key14306015
Report Number8030647-2022-00131
Device Sequence Number20090276
Product Code BSY
UDI-Device Identifier00609038938349
UDI-Public00609038938349
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/08/2026
Device Model Number210
Device Catalogue NumberN/A
Device Lot Number30154260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2022
Supplement Date Received by Manufacturer05/17/2022
09/20/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date06/01/2022
09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-