BOSTON SCIENTIFIC CORPORATION ACCUTRAC; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M0068404110
Medical Device Problem Codes Break (1069); Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Investigation results: the returned accutrac laser fiber was analyzed, and a visual evaluation noted that the laser fiber was returned in one piece.The fiber body measures 318.2 cm from the end of the connector to the end of the fiber body.The exposed glass tip measured 4 mm and was in good condition.A functional evaluation revealed that the light from the sma connector was distorted, indicating a fracture within the fiber connector.No other problems with the device were noted.The reported event of the accutrac laser fiber failed to emit energy was confirmed.It is likely that procedural handling of the device during set-up or use contributed to the fracture within the fiber connector.A broken fiber within the connector of the device can result in insufficient aiming beam and low laser energy output, up to a complete inability for the fiber to fire.Review and analysis of all available information indicated that the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that an accutrac laser fiber was used during a ureteroscopy procedure in the ureter, performed on (b)(6) 2022.The laser unit used was a 100h laser console.During the procedure, the laser fiber did not produce laser energy.The procedure was completed with another accutrac laser fiber.There were no patient complications reported as a result of this event.Note: this event has been deemed reportable based on the investigation finding of laser fiber broken within the connector.Please see block for full investigation details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUTRAC
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
COHERENT INCORPORATED
5100 patrick henry drive
santa clara CA 95054
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14306544
Report Number3005099803-2022-02518
Device Sequence Number12018417
Product Code GEX
UDI-Device Identifier08714729802716
UDI-Public08714729802716
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K110686
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/09/2024
Device Model NumberM0068404110
Device Catalogue Number840-411
Device Lot Number0000011547
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/15/2022
Initial Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-