• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 BD INTRAOSSEOUS NEEDLE SET KIT; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 BD INTRAOSSEOUS NEEDLE SET KIT; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number N/A
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of 127808 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "the luer nut on the extension set is very "sticky" and tough to adjust and attach to the needle." no other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a ¿sticky¿ luer adapter was confirmed; however, the root cause was not identified.The product returned for evaluation was one intraosseous (io) needle extension set.Light usage residues were observed within the male luer adapter.Attachment of the extension set to non-complainant accessories was successful and unremarkable.Attempts to infuse water through the sample using a 12ml syringe revealed the sample to be patent to infusion and aspiration with no observed leaks.No leaks were observed at the connection points with non-complainant accessories.Tactile inspection of the sample revealed resistance when manipulating the male luer adapter locking collar.Microscopic inspection of the sample confirmed white crystalline residue within the male luer adapter.Some resistance was encountered when manipulating the male luer adapter locking collar; however, the exact root cause of that resistance was not identified.It appeared that residue accumulation within the luer adapter may have contributed.H3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported "the luer nut on the extension set is very "sticky" and tough to adjust and attach to the needle." no other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTRAOSSEOUS NEEDLE SET KIT
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14307596
Report Number3006260740-2022-01632
Device Sequence Number20077592
Product Code FMI
UDI-Device Identifier00801741163609
UDI-Public(01)00801741163609
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K203193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberD015451NK
Device Lot Number127808
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/19/2022
Supplement Date Received by Manufacturer06/08/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date08/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-