• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH CMN FEMORAL NAIL, CCD 130°, RIGHT, 11.5 MM, 34 CM; ZIMMER NATURAL NAIL SYSTEM CEPHALOMEDULLARY NAILS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH CMN FEMORAL NAIL, CCD 130°, RIGHT, 11.5 MM, 34 CM; ZIMMER NATURAL NAIL SYSTEM CEPHALOMEDULLARY NAILS Back to Search Results
Model Number N/A
Medical Device Problem Code Human-Device Interface Problem (2949)
Health Effect - Clinical Codes Fall (1848); Bone Fracture(s) (1870); Insufficient Information (4580)
Date of Event 03/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The manufacturer did not receive x-rays, or other source documents for review.The manufacturer did not receive the device for investigation.The lot number of the device was received.The device history records will be reviewed during investigation.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available an updated report will be submitted.Zimmer biomet¿s reference number of this file is (b)(4).
 
Event or Problem Description
It was reported that: explant of the nail after the patient fell on the floor.Patient outcome - revision.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The product was not returned for investigation.Therefore, visual and dimensional evaluation could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.The raw material certificate was reviewed with no anomalies noted.It was reported that the patient had a fall event prior to the revision.If and to what extent the fall event caused or contributed to the revision surgery remains unknown.It can be assumed that due to the fall event, the patient suffered an implant fracture and a re-fracture of the bone.However, due to a lack of information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CMN FEMORAL NAIL, CCD 130°, RIGHT, 11.5 MM, 34 CM
Common Device Name
ZIMMER NATURAL NAIL SYSTEM CEPHALOMEDULLARY NAILS
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14307852
Report Number0009613350-2022-00271
Device Sequence Number5223736
Product Code HSB
UDI-Device Identifier00889024299504
UDI-Public00889024299504
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K091566
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number47-2493-342-11
Device Lot Number3020706
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date08/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/19/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-