Within 3 minutes of initiating crrt (continuous renal replacement therapy) on a patient the tubing was saturated with air, leaving a large amount of air bubbles in the line.This set was unusable, and the blood could not be returned to the patient.The rn secured another crrt prismax machine and hung another set from the same supply room.Within 2 minutes of initiating the therapy, this set was also filled with diffuse air and the blood could not be returned to the patient.Upon getting another set from a different batch off unit, the therapy was able to be initiated without difficulty.Manufacturer response for crrt tubing, (brand not provided) (per site reporter): notified baxter.[redacted date]: baxter complaint # (b)(4); responded to baxter questionnaire.[redacted date]: rma received.[redacted date]: sample shipped ups.
|