Model Number 11403002
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
During an evlt procedure, the gold tip of the nevertouch laser fiber came loose and detached from the fiber when the physician was wiping off before reinserting into the patient.This occurred outside of the patient; therefore, there was no harm experienced.The procedure was completed with another same device.
 
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.However the customer provided a photo.The photo shows the gold tip is detached from the fiber, noted there is what appears to be dried blood on the gold tip.Examination of the photograph noted the gold tip tube is completely separated from the fiber assembly, there is a noticeable dark area on one end of the gold tip tube, and the distal end of the fiber appears to be very clean (no thermal degradation).The customer's reported complaint description of tip detached is confirmed via picture provided by end user, however, the fiber was not returned for evaluation.Without receiving a complaint sample for evaluation a definitive root cause cannot be determined.A possible root cause for the event could have been excessice force used when withdrawing the fiber through the tre-sheath, causing the tip tube bond to allow seperation form the fiber during the cleaning process.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the instructions for use which is supplied to the end user with this catalog number contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing, or excessive bending.Do not coil the fiber tighter than a diameter of 20cm" and "pull the sheath back and lock it to the sheath-lok fitting.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENACURE NEVERTOUCH FIBER
Common Device Name
VENACURE ENDOVENOUS LASER TREATMENT FIBER
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key14309870
Report Number1319211-2022-00018
Device Sequence Number17471193
Product Code GEX
UDI-Device IdentifierH787114030020
UDI-PublicH787114030020
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K171921
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number11403002
Device Catalogue Number11403002
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2022
Initial Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-