| Model Number |
21-7624-24 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No lot number was provided; therefore, device history record review could not be completed.A product sample was received for evaluation.Visual and functional testing were performed.The samples returned were tested using the hydrostatic vessel.A leak was detected fleeing from the joint of the pump tube and "tube 105".The most probable root cause of the reported event was found to be lack/insufficient cyclohexanone adhesive".
|
| |
|
Event or Problem Description
|
|
It was reported that fluid leaked from the connector.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
While performing a review of submitted mdr report it was discovered that this file was inadvertently assessed as reportable.No patient death, serious injury, and no evidence of a reportable product malfunction.The hazardous situation for the given complaint device would not likely cause or contribute to a death or serious injury if the malfunction were to recur.File (b)(4) is no longer considered reportable, please disregard any mdr reports associated with it.
|
| |
|