| Model Number |
CD3265-40Q |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted.The patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Premature battery depletion was confirmed by analysis.A device evaluation was performed, and no sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
|
| |
|
Search Alerts/Recalls
|