ABBOTT PROMOTE+ CRT-D W/SJ4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3211-36Q
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
During a clinic follow-up, a decrease in the battery longevity was noted on the device and it was suspected that there was an overestimation of the device longevity at a previous follow-up.Technical support was contacted and the device was explanted due to normal elective replacement indicator (eri) and replaced.The patient was stable and will continue to be monitored.
 
Additional Manufacturer Narrative
The reported event of overestimation was confirmed.The device was above elective replacement indicator (eri) upon receipt.Longevity analysis revealed the battery depletion to be normal.The device functioned normally during analysis and no anomalies were revealed.The longevity overestimation was determined to be due to an estimation inaccuracy in the merlin programmer remaining longevity calculation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMOTE+ CRT-D W/SJ4 CONNECTOR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14310572
Report Number2017865-2022-09483
Device Sequence Number12106594
Product Code NIK
UDI-Device Identifier05414734503679
UDI-Public05414734503679
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2016
Device Model NumberCD3211-36Q
Device Catalogue NumberCD3211-36Q
Device Lot NumberS000007249
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2022
Supplement Date Received by Manufacturer06/15/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-