Model Number 20188
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr.: the express sd was returned to boston scientific for analysis.The device was inspected for any damage or irregularities.The express sd was returned with a partial unknown guidewire still inserted.Between the guidewire and catheter was a small piece of what looked like a white stretchy material similar to nitrile gloves.The device showed no signs of saline, but no damage was noted on the device.
 
Event or Problem Description
It was reported that device contamination occurred.A 5.0mmx15mmx150cm express sd stent was selected for use.However, it was noted that the sterile scrubbed glove was torn and caught onto the stent upon removing it from the hoop of the sterile packaging.So the stent had to be removed from the field, the contaminated gloves were removed and replaced with a new one.The procedure was completed with another of the same device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS SD RENAL/BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14310680
Report Number2134265-2022-04405
Device Sequence Number5286835
Product Code NIN
UDI-Device Identifier08714729390398
UDI-Public08714729390398
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K152607
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/18/2024
Device Model Number20188
Device Catalogue Number20188
Device Lot Number0028226466
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/11/2022
Initial Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age83 YR
Patient SexFemale
Patient RaceWhite
-
-