Model Number 21-7321-24
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the luer connector was defective.No patient injury was reported.
 
Additional Manufacturer Narrative
No lot number was provided; therefore, device history record review could not be completed.A product sample was received for evaluation.Visual and functional testing were performed.A review of the manufacturing process for was conducted by quality engineer in order to verify that there are no situations or practices that could create the event.No discrepancies were found.The sample was visually inspected at a distance of 12 to 16 under normal conditions of illumination.The asv from the sample received was compared with other piece from the same component and it was found that the nozzles from the valve is bigger than the sample taken from production floor.The asv was measured using a caliper and it was detected that the component is with off specification.The most probable root cause is that the asv was received defective from the supplier.A supplier notification was issued in order to knowledge the vendor that the part was involved in a complaint.
 
Additional Manufacturer Narrative
Other, other text: while performing a review of submitted mdr report it was discovered that this file was inadvertently assessed as reportable.No patient death, serious injury, and no evidence of a reportable product malfunction.The hazardous situation for the given complaint device would not likely cause or contribute to a death or serious injury if the malfunction were to recur.Report 3012307300-2022-07835 is no longer considered reportable, please disregard any mdr reports associated with it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD MEDICATION CASSETTE RESERVOIRS WITH FLOW STOP
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14310762
Report Number3012307300-2022-07835
Device Sequence Number5313352
Product Code FPA
UDI-Device Identifier10610586029639
UDI-Public10610586029639
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
K040636
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Pharmacist
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number21-7321-24
Device Catalogue Number21-7321-24
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2022
Supplement Date Received by Manufacturer06/07/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0148-2022
Patient Sequence Number1
Patient SexUnknown
-
-