• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INNOVA VASCULAR; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INNOVA VASCULAR; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number 26924
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that inadvertent deployment occurred.A 5mmx 80mm x 130cm innova vascular stent was selected for a superficial femoral artery intervention procedure.Upon removing the innova vascular stent from the packing hoop, it was observed that the stent was deployed outside of the deployment shaft.The device was replaced with another of the same device to complete the procedure.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device analysis by manufacturer: returned product consisted of an innova self-expanding stent system.The outer sheath, tip, inner sheath, and the remainder of the device were checked for damage.Visual examination revealed the thumbwheel lock in place and the pull handle in place.The flushing luer was not attached or returned.Microscopic examination revealed the stent had deployed 2.3 cm from the outer sheath.A kink was seen 5 mm from the nosecone.
 
Event or Problem Description
It was reported that inadvertent deployment occurred.A 5mmx 80mm x 130cm innova vascular stent was selected for a superficial femoral artery intervention procedure.Upon removing the innova vascular stent from the packing hoop, it was observed that the stent was deployed outside of the deployment shaft.The device was replaced with another of the same device to complete the procedure.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA VASCULAR
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14310787
Report Number2134265-2022-05087
Device Sequence Number20162057
Product Code NIP
UDI-Device Identifier08714729873921
UDI-Public08714729873921
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P140028
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/09/2026
Device Model Number26924
Device Catalogue Number26924
Device Lot Number0026944570
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/13/2022
Supplement Date Received by Manufacturer05/12/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date06/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
-
-