It was reported that on an unknown date, during inspection of the back up devices it was discovered that the gold peg of one (1) recon locking aiming arm for lfn nails-ex has snapped off, the threads of one (1) extraction screw for ti femoral & tibial nails and one (1) extraction screw/m8 for suprapatellar are stripped, and one (1) depth gauge for 2.0/2.4 and 2.7 screws was stuck in place.There was no patient involvement.Upon manufacturer investigation, the device was found to be discolored from the ball handle and deformed from the shaft.This report is for a reaming rod push rod with ball handle.This is report 6 of 6 for (b)(4).
|
|
Reporter is a synthes employee.A manufacturing related potential cause was not suspected, therefore, per franchise complaint product investigation procedure, no manufacturing record evaluation is required.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the reaming rod push rod with ball handle was found to be discolored from the ball handle and deformed from the shaft.Signs of rust were not visible.A dimensional inspection for the reaming rod push rod with ball handle was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the reaming rod push rod with ball handle would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed the following drawings reflecting the current and manufactured revisions were reviewed: reaming rod push rod with ball handle (current version).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|