Model Number 21-7081-24
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record (dhr) review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.A product sample was received for evaluation.Visual and functional testing were performed.The sample received was visually inspected and no damage nor workmanship defect were detected.Leak testing on the sample received was performed to look for unusual functions.No leaks were detected fleeing from the inj.Site port nor other site.A review of the manufacturing process for was conducted by quality engineer in order to verify that there are no situations or practices that could create the event.No discrepancies were found.The most probable root cause are: the inj.Site was received damaged from the supplier.The inj.Site became damage after the product left the manufacturing facilities.A supplier notification was issued in order to knowledge the vendor that the part was involved in a complaint.
 
Event or Problem Description
It was reported that a part of the septum was defective.No patient injury was reported.
 
Additional Manufacturer Narrative
Other, other text: while performing a review of submitted mdr report it was discovered that this file was inadvertently assessed as reportable.No patient death, serious injury, and no evidence of a reportable product malfunction.The hazardous situation for the given complaint device would not likely cause or contribute to a death or serious injury if the malfunction were to recur.Report 3012307300-2022-07844 is no longer considered reportable, please disregard any mdr reports associated with it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD ADMINISTRATION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14310982
Report Number3012307300-2022-07844
Device Sequence Number14796552
Product Code FPA
UDI-Device Identifier10610586021480
UDI-Public10610586021480
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K991599
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Pharmacist
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/26/2024
Device Model Number21-7081-24
Device Catalogue Number21-7081-24
Device Lot Number3780633
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2022
Supplement Date Received by Manufacturer06/07/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/15/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0148-2022
Patient Sequence Number1
Patient SexUnknown
-
-