| Model Number |
26055CB |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
04/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As per manufacturer incident report #(b)(4) received from the factory in (b)(4): the event occurred in (b)(6).According to the information available, the ceramic tip of the 26055cb broke in the patient's uterus.A few of the broken parts are still in the patient's uterus.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected device was requested for further investigation, but has not returned yet.
|
| |
|
Additional Manufacturer Narrative
|
|
The instrument has dents on the shaft.These occur when the inner shaft is not pulled straight out of the outer shaft.Pulling out the inner shaft at an angle exerts pressure on the ceramic insert, causing it to break.It is already included in the ifu that there is a risk of injury if the device is overloaded.
|
| |
|