Model Number 10-0000-07
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
In review of all associated customer-provided documentation and internally generated investigation notes against the criteria set forth in document (b)(4) (medical device reporting standards operating procedure), this complaint does meet the requirements for an mdr in the countries where this product is sold or distributed.Filing individual mdr out of abundance of caution due to the remote possibility of a false result contributing to a death or serious injury.
 
Event or Problem Description
The service visit at katherine (b)(6) hospital.Was scheduled per capa-21-014 within work order wo-00199948.During the site visit, the elution heater required recalibration due to failing initial verification at setpoints 1 and 2.Field quality updated risk assessment question 2 to 'yes' since the setpoint temperatures failed outside the accepted limits established by the verigene i assay temperature tolerance guard band risk report ((b)(4)).Per the verigene assay temperature tolerance (guard-band) study report ((b)(4)), the study evaluated the assay performance of verigene® rp flex, ep, cdf, bc-gp and bc-gn assays at varying critical temperature parameters and established a functional range of temperature for assay tolerance above and below the nominal set points.It was determined within (b)(4), that the elution heater has a lower tolerance of -4degree celsius and an upper tolerance of greater than or equal to 5degree from the setpoint temperature 1 (75.0c) and setpoint 2 (85.0c).In reviewing the attached naplogs, setpoint 1 for the elution heater had an adjusted temperature of +86.1c and setpoint 2 had an adjusted temperature of +97.9c; both higher than the acceptable range per (b)(4).Due the finding, risk assessment question 2 was updated to 'yes'.The customer site is currently running bc-gp, bc-gn, and ep verigene assays.The elution heater verification failure can potentially impact assay performance for all three assays ran at the customer site per (b)(4).Within an 18-month timeframe from the verification failure, the customer site has not reported any patient sample false results on either assay.The customer site reported no false results for the sp processor s/n (b)(4).There is no indication of this verification failure resulting in an adverse effect to patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERIGENE PROCESSOR SP
Common Device Name
VERIGENE PROCESSOR SP
Manufacturer (Section D)
LUMINEX CORPORATION
12212 technology blvd
austin TX 78727
Manufacturer Contact
jennifer svoboda
12212 technology blvd
austin, TX 78727
MDR Report Key14311544
Report Number1650733-2022-00021
Device Sequence Number4677059
Product Code NSU
UDI-Device Identifier00840487101483
UDI-Public00840487101483
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number10-0000-07
Device Catalogue Number10-0000-07
Device Lot Number18128002
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2022
Initial Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/08/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-