Model Number 68LP-525
Medical Device Problem Codes Material Rupture (1546); Device Handling Problem (3265)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Healthcare professional reported right side rupture.Device remains implanted.
 
Event or Problem Description
Additionally, ¿no.68 low profiles 525cc filled to 400cc previously¿ was reported.The device was explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified fold creases, wear abrasion, one extended opening and yellow particles in device inner surface.A microscopic analysis and leak test were performed which identified one opening crease smooth.Fill inspection was performed and identified no blockage in the valve.Based on the device analysis the final assessment is: one opening crease smooth in the side anterior assessed as fold flaw opening.Additional, changed, and/or corrected data.
 
Event or Problem Description
Healthcare professional reported "right side rupture." later, healthcare reported "filled to 400cc (for a 575 implant)" and "any creases." the device has been explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 68 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key14311578
Report Number9617229-2022-07336
Device Sequence Number14780786
Product Code FWM
UDI-Device Identifier10888628000650
UDI-Public10888628000650
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number68LP-525
Device Catalogue Number68LP-525
Device Lot Number3044185
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/14/2022
Supplement Date Received by Manufacturer05/11/2022
06/17/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date06/09/2022
07/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age27 YR
Patient SexFemale
Patient Weight86 KG
-
-