Catalog Number 74121138
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 01/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that on literature review "metal-on-metal hip resurfacing arthroplasty provides excellent long-term survivorship and function in patients with a good sized femoral head.", 1 patient suffered femoral neck fracture 2.33 years after a bhr and required a revision surgery.There's no more information available.
 
Additional Manufacturer Narrative
H3, h6: it was reported that the patient suffered femoral neck fracture and required a revision surgery.There is no more information available and the devices, used in treatment, were not available for analysis.As no device batch numbers were provided for investigation manufacturing record review, device labelling / ifu review could not be performed.If more information is received, this investigation will be reopened.Without a definitive batch number, a complete review of the historical complaints data cannot be performed for the femoral head.A review of historical complaints data was performed using part numbers and the reported failure modes to evaluate patterns of repeated failures or defects in a timeframe prior 12 months as of the complaint aware date.No other similar complaints were identified for the part number and the reported/related failure mode.As the device is no longer sold, no action is to be taken.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions required.Without clinically relevant patient-specific supporting documentation, a thorough medical investigation could not be performed.The images provided in the article have been interpreted within the text; therefore, no further analysis of the images is required.The root cause and/or patient outcome beyond that which was documented in the article could not be confirmed nor concluded; therefore, no further medical assessment is warranted at this time.Based on the information provided we cannot confirm or further investigate the reported complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESURFACING FEMORAL HEAD 38MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14311950
Report Number3005975929-2022-00202
Device Sequence Number5026302
Product Code NXT
UDI-Device Identifier03596010502766
UDI-Public3596010502766
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number74121138
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2022
Supplement Date Received by Manufacturer09/15/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2745-2015
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
Patient SexFemale
-
-