Model Number CTB73
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of the investigation.
 
Event or Problem Description
Procedure performed: right laparoscopic colectomy.Event description: [translation]: when inserting a laparo device into the adapted trocar, a piece of rubber from the membrane got broken and fell (by chance in the operator's field of vision) into the patient's abdominal cavity and could have been unseen and left in the abdomen.Additional information received by email from applied medical representative on (b)(6) 2022.Name of procedure being performed: right laparoscopic colectomy.Patient status : nothing to report due to the immediate reaction of the team.Pieces fell into the patient and were all retrieved from the patient.The inner membrane detached from the seal.The component was black.The entire component appeared to be torn off.Self-suturing pliers diam 12 were being used at the time of the incident.Type of intervention: pieces fell into the patient and were all retrieved from the patient.Patient status: nothing to report due to the immediate reaction of the team.
 
Additional Manufacturer Narrative
A portion of the event unit was returned to applied medical for evaluation, along with a black fragment.Visual inspection of the event unit confirmed the complainant¿s experience of seal component separation.The black fragment matched the missing portion of the septum, an internal rubber component of the seal.Based on the condition of the returned unit, it is likely that the reported event was caused by non-axial insertion or removal of asymmetrical instruments through the trocar.Applied medical¿s instructions for use (ifu) states, "extra care should be used when inserting angular and asymmetrical instruments, such as 'j' hooks and clip appliers.All instruments should be centered axially when inserted through the seal to prevent tearing." this event was initially reported based on the description of the event.However, based on the evaluation of the event unit and additional information, applied medical determined that this event is not reportable as it is unlikely to cause or contribute to death or serious injury.
 
Event or Problem Description
Procedure performed: right laparoscopic colectomy.Event description: [translation]: when inserting a laparo device into the adapted trocar, a piece of rubber from the membrane got broken and fell (by chance in the operator's field of vision) into the patient's abdominal cavity and could have been unseen and left in the abdomen.Additional information received by email from applied medical representative on 13apr22: name of procedure being performed: right laparoscopic colectomy.Patient status : nothing to report due to the immediate reaction of the team.Pieces fell into the patient and were all retrieved from the patient.The inner membrane detached from the seal.The component was black.The entire component appeared to be torn off.Self-suturing pliers diam 12 were being used at the time of the incident.Additional information received by email from applied medical representative on 16june22: as discussed by phone, based on the device returned and initial description of the incident the customer experience a seal component fragmentation and not a seal component dislodgment.Type of intervention: pieces fell into the patient and were all retrieved from the patient.Patient status: nothing to report due to the immediate reaction of the team.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CTB73, 12X100 KII SH BLD ZTHR 6/BX
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key14312121
Report Number2027111-2022-00589
Device Sequence Number17399986
Product Code GCJ
UDI-Device Identifier00607915123727
UDI-Public(01)00607915123727(17)240310(30)01(10)1408624
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K072674
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/10/2024
Device Model NumberCTB73
Device Catalogue Number101472289
Device Lot Number1408624
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2022
Supplement Date Received by Manufacturer04/13/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date06/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SELF-SUTURING PLIERS DIAM 12
Patient SexPrefer Not To Disclose
-
-