Model Number CW-USLTA
Medical Device Problem Code Fire (1245)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when the cw-uslta was activated with the cyberwand generator, the cable caught fire and flames shot out of the end of the handpiece.Customer site facility biomed confirmed that the handpiece was not in the patient at the time of the incident, the user had the handpiece in his hand testing the handpiece.The issue occurred during an unspecified procedure.The intended procedure according to the customer completed with another device.Customer site biomed confirmed that the fire reported to have come from the transducer tip.It was not from the generator and there was no patient injury.There was no patient harm or injury reported due to the event.No user injury reported.This report is related to a report with patient identifier (b)(6) (generator).
 
Additional Manufacturer Narrative
The subject device was not returned for evaluation.In a follow up communication with the customer, it was conveyed that the device is not returning.Customer stated they did not discarded the device but kept it in the facility.The customer stated that the device cable was burned and cannot be tested and therefore they will not return for evaluation.Supplemental report(s) will be submitted should any relevant new information is available and or received.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.As the device did not return to olympus, a physical device inspection could not be performed.Further, the serial number provided was not in the system, a device history record review could not be performed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the investigation results, and since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined.The following information is stated in the instructions for use: "never attempt to service the device when it is connected to a power source.Hazardous voltages inside the device may cause severe electrical shock.Disconnect the power cord before servicing." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CW-USLTA CYBERWAND ADVANCED TRANSDUCER
Common Device Name
LITHOTRIPTOR, ELECTRO-HYDRAULIC
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough PA 01772
Manufacturer Contact
brian motter
9600 louisiana avenue north
brooklyn park, MN 55445
4848965250
MDR Report Key14312262
Report Number3011050570-2022-00053
Device Sequence Number14702432
Product Code FFK
UDI-Device IdentifierH891CWUSLTA2
UDI-PublicH891CWUSLTA2
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K120303
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCW-USLTA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2022
Supplement Date Received by Manufacturer05/10/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR
Patient SexUnknown
-
-