| Model Number |
DRX-Revolution Mobile X-ray System |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
Bruise/Contusion (1754)
|
| Date of Event |
04/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2022, carestream health (csh) was informed of an incident related to the drx-revolution mobile x-ray system which occurred at south east health- us.The site reported that a rad tech was backing the revolution plus portable out of the patient room and the unit suddenly backed up pinning her arm against the wall.Her arm was sore and bruised.The tech went to the emergency room for an x-ray of her elbow which showed no broken bones.The injury consisted of a sore and bruised arm.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This investigation was conducted by the dealer, service provider in conjunction with carestream support.Based on the investigation and facts presented by the dealer fe, it is unclear what the problem or root cause was in this case as the dealer could not provide additional details regarding the issue or how the tech was driving the system at the time of the incident.The dealer replaced several components related to the drive system: strain gauges, multi access board and performed cart calibrations which provided a smoother drive; however, it is unknown if these were warranted as he could not reproduce the initial reported scenario at any point in time during the investigation.The system is driving without any issues and is currently in use by the site.There are no further actions to be taken.Carestream will monitor for these types of incidents.
|
| |
|