ARROW INTERNATIONAL LLC ARROW CVC KIT: 3-LUMEN 12 FR X 20 CM; CATHETER INTRAVASCULAR THERAPE Back to Search Results
Model Number IPN036855
Medical Device Problem Codes Unraveled Material (1664); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The mdr report key/report number is (b)(4).The mdr text key is (b)(4).
 
Event or Problem Description
Complaint found in maude database reports: "the advanced practice provider (app) was placing a line for continuous renal replacement therapy (crrt).Ultrasound was used to find vessel, the guide wire was placed, needle removed, small incision made, and line was placed.While removing the guide wire it became what felt like "stuck".Upon removal the "hooked" part of the guide wire was frayed.".
 
Additional Manufacturer Narrative
Qn(b)(4).The mdr report key/report number is (b)(4).The mdr text key is (b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
Complaint found in maude database reports: "the advanced practice provider (app) was placing a line for continuous renal replacement therapy (crrt).Ultrasound was used to find vessel, the guide wire was placed, needle removed, small incision made, and line was placed.While removing the guide wire it became what felt like "stuck".Upon removal the "hooked" part of the guide wire was frayed.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC KIT: 3-LUMEN 12 FR X 20 CM
Common Device Name
CATHETER INTRAVASCULAR THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key14312630
Report Number9680794-2022-00273
Device Sequence Number17399994
Product Code FOZ
UDI-Device Identifier60801902122089
UDI-Public60801902122089
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K970864
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model NumberIPN036855
Device Catalogue NumberAK-15123-F
Device Lot Number13F21J0523
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2022
Supplement Date Received by Manufacturer06/02/2022
Initial Report FDA Received Date05/06/2022
Supplement Report FDA Received Date06/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-